An electrostatic in-line charge-state purification system for multicharged ions in the kiloelectronvolt energy range

Anna Lévy

Résumé

The performance of a newly built omega type electrostatic analyzer designed to act as an in-line charge-state purification system for ions in the kiloelectronvolt energy range is reported. The analyzer consists of a set of four consecutive electrostatic 140° concentric cylindrical electrodes enclosed by Matsuda electrodes. This setup was recently tested and validated using O5+, Ar9+, and Xe20+ ion beams at an energy of 14 qkeV at the ARIBE facility. A resolving power of 10.5 and a transmission of 100% of the desired charge state are measured allowing a good purification of incoming ion beams with charge states up to 10+ and a fairly good purification for charge states at least up to 20+. In comparison with other in-line solutions such as the Wien filter, our system has the advantage of being purely electrostatic and therefore lacking common drawbacks as, for example, hysteresis.
Fichier principal
Vignette du fichier
FISIC_Revised.pdf (1.32 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-02269117 , version 1 (04-01-2021)

Identifiants

Citer

Daniel Schury, Ajit Kumar, Alain Méry, Jean-Yves Chesnel, Anna Lévy, et al.. An electrostatic in-line charge-state purification system for multicharged ions in the kiloelectronvolt energy range. Review of Scientific Instruments, 2019, 90 (8), pp.083306. ⟨10.1063/1.5093407⟩. ⟨hal-02269117⟩
165 Consultations
93 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More