On the characterisation of a Bragg spectrometer with X-rays from an ECR source

Résumé

Narrow X-ray lines from helium-like argon emitted from a dedicated ECR source have been used to determine the response function of a Bragg crystal spectrometer equipped with large area spherically bent silicon (111) or quartz (10$\bar{1}$) crystals. The measured spectra are compared with simulated ones created by a ray-tracing code based on the expected theoretical crystal's rocking curve and the geometry of the experimental set-up.
Fichier principal
Vignette du fichier
NIM_BAR-04_132_nimnorm.pdf (431.85 Ko) Télécharger le fichier

Dates et versions

hal-00002575 , version 1 (17-08-2004)
hal-00002575 , version 2 (22-12-2004)

Identifiants

Citer

D.F. Anagnostopoulos, S. Biri, D. Gotta, A. Gruber, Paul Indelicato, et al.. On the characterisation of a Bragg spectrometer with X-rays from an ECR source. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2005, 545, pp.217. ⟨10.1016/j.nima.2005.01.311⟩. ⟨hal-00002575v2⟩